Parkinson’s disease motor symptoms often begin in one upper limb, and tremor and bradykinesia are the earliest and most readily observed of them. A rating scale can assign a grade from 0 to 4, but it struggles to capture subtle changes before and after a single dose of medication. For engineers building wearable devices and rehabilitation assessment systems, the question is not whether movement can be measured, but how the raw waveform of a 6-axis IMU sensor can be turned into quantitative metrics that clinicians are willing to trust.
1. Beyond Rating Scales: Why a Wearable IMU Is Needed
Scale-based assessment has three main shortcomings. First, accuracy is poor. Scale assessment is not fully objective; it depends on the clinician’s subjective experience, and inter-rater agreement is low. Second, resolution is low. Taking the most widely used UPDRS as an example, each item can only be assigned one of five integer levels from 0 to 4. It can detect obvious symptom improvement or worsening, but it is powerless against small changes, making the effect of short-term medication difficult to quantify. Third, timeliness is poor. Clinical assessment addresses the patient’s current state, basically within a hospital department; the patient may not show symptoms at the time of the visit, and the scale score cannot accurately reflect the severity of motor impairment in the patient’s actual life.
An ideal assessment method should provide objective, quantitative, and timely information about the patient’s symptoms to help clinicians diagnose and improve treatment plans.
Optical motion capture is currently the most authoritative motion capture method. However, optical motion capture involves data processing steps including recognition, tracking, and coordinate calculation; its computational workload is large, the related equipment is expensive, optical conditions are demanding, and capture performance is poor under occlusion. MEMS inertial motion capture equipment can overcome the insufficient measurement accuracy of optical systems in low-light environments and under motion occlusion, and offers low equipment cost, relatively high measurement accuracy, fast response, and suitability for real-time motion capture, while supporting parallel capture of multiple subjects’ motion states.
An inertial measurement unit that integrates accelerometers and gyroscopes on a single chip has been widely used in human motion analysis because of its low power consumption and wearability. These advantages make a wearable IMU sensor very suitable for long-term patient monitoring and mobile assessment of PD patients in the home environment, making it convenient to compare medication effects and fluctuation phenomena and to help clinicians improve treatment plans.

2. Defining the Measurement Target: Tremor and Bradykinesia
Tremor is a typical symptom of PD. It refers to involuntary movement of rhythmic oscillation of a body segment, mostly beginning in the early stage of PD, and is divided into resting tremor, postural tremor, and action tremor. Among these, the most common in PD patients is the “pill-rolling” resting tremor of the thumb and index finger; about 75% of PD patients have this symptom, and it is one of the most characteristic clinical symptoms of PD. Neuromedical research shows that the clinical feature of PD tremor is a frequency range of 4–7 Hz, with tremor frequency decreasing over time and tremor amplitude negatively correlated with frequency.
Bradykinesia is another major symptom of PD. It is a hallmark of basal ganglia disorder, mostly occurring in the middle and late stages, and includes difficulty in planning, initiating, and executing movement, or in performing continuous and simultaneous tasks. It manifests clinically as slowness in daily activities and reaction time, especially in tasks requiring fine motor control.
3. Data Acquisition: Where to Place Nodes, and How to Set Range and Sampling Rate
Specifically, the design should meet the following requirements: (1) the size and weight of devices worn on limb segments should be as small as possible to avoid affecting human movement; (2) wireless transmission should be used between devices as far as possible to avoid excessive wires affecting human movement; (3) system power consumption should be as low as possible, ensuring at least 5 hours of continuous operation.
The final design consists of two subsystems: a finger node and a body node. The finger node subsystem includes 2 acquisition modules for collecting movement signals from the fingers of the left and right hands, with 2 inertial sensors connected at the front and placed on the thumb and index finger with finger cots. The body node subsystem includes 10 acquisition modules, each with 1 inertial sensor. Of these, 8 modules are distributed on the left and right forearms, upper arms, thighs, and lower legs, and the remaining 2 modules are placed on the upper back and lower back.

The acquisition module integrates a 3-axis accelerometer and a 3-axis gyroscope on a single chip, forming a 6-axis IMU sensor that provides 6-DOF motion information. Through programming, the accelerometer range is set to ±8 g and the gyroscope range to ±1000°/s. All modules use a uniform 100 Hz sampling rate. According to the Shannon sampling theorem, for human motion signals with frequencies not exceeding 20 Hz, a 100 Hz sampling rate is sufficient.

4. Preprocessing: Separate Voluntary Movement from Tremor
Signal preprocessing generally uses a classic Butterworth IIR filter. When tremor signals need to be preserved, the raw signal is high-pass filtered with a cutoff frequency of 1 Hz to remove voluntary movement and the gravity component of the acceleration signal. When voluntary movement needs to be preserved, the raw signal is high-pass filtered with a cutoff frequency of 0.25 Hz to remove low-frequency signal drift and the gravity component of the acceleration signal.
The traditional method is based on a classic linear filter. Its theoretical basis, the Fourier transform, requires the signal to be linear and stationary; however, this assumption is difficult to satisfy in actual physiological signals, and human motion signals are typical nonlinear, non-stationary signals. Moreover, tremor and finger movement signals may partially overlap in bandwidth, making it difficult for a linear filter to completely remove the tremor component.
4.1 Separating Intermittent Tremor with HHT
HHT is a new nonlinear, non-stationary signal processing method consisting of two steps: empirical mode decomposition (EMD) and Hilbert spectral analysis. First, the finger opening-closing movement signal acquired by the inertial sensor is decomposed by EMD; then, based on prior knowledge, a modal selection algorithm is designed, and the selected modes are used to reconstruct the desired movement signal.
After EMD, Why Modal Selection Still Matters
The empirical nature of HHT and its ability to handle nonlinear and non-stationary signals allow it to adaptively identify faster or slower dynamic processes on the same time scale, thereby selectively preserving the local characteristics of the original time series. This is very important for signals in which tremor components appear intermittently.
The HHT filtering algorithm achieves a higher SNR and lower noise RMSE than the LPF.
5. Feature Engineering: Turning Waveforms into Interpretable Metrics
5.1 Tremor: Amplitude and Band Power
Two types of feature parameters are calculated for the 6-axis signal of each task: time-domain variables include the root mean square (RMS) of acceleration and angular velocity; frequency-domain variables include the peak amplitude of the power spectrum (ppeak), the frequency corresponding to the peak amplitude (fpeak), the power in the tremor band from 4 to 7 Hz (pt), and its percentage of the total power spectrum power (Ratio). According to related research, parameters related to tremor amplitude increase exponentially with severity level, and after taking the logarithm they are basically linearly related to severity level.
5.2 Bradykinesia: Speed, Amplitude, and Rhythm Features
To quantify bradykinesia as comprehensively as possible, medicine examines three aspects: slow movement speed, reduced motor function, and rhythm disorder. Following these three aspects, 15 time-domain and frequency-domain parameters are extracted from the finger opening-closing and grasping signals.
5.3 Rhythm: Approximate Entropy and Sample Entropy
Two nonlinear indicators for measuring the regularity of time series are approximate entropy (ApEn) and sample entropy (SampEn). For clinical data, when the sequence length N is about 2000, both statistics show strong stability; the template length m is chosen as 2, and the threshold r is set to 0.1–0.25 times the standard deviation of the sequence, with a typical value of 0.2.
6. From Features to Grading: PCA and Multi-class SVM
A one-versus-one method is used to train a multi-class SVM, with each classifier outputting 5 classes labeled 0–4, representing symptom severity from low to high. Parameters are selected by grid search, and the SVM model is trained with 5-fold cross-validation to select the parameter set with the highest accuracy.
For resting tremor, the overall accuracy using the original feature set is 0.913, and the overall accuracy after PCA dimensionality reduction is 0.898; classification errors mainly occur at scores of 0 and 1, while accuracy is above 0.95 for scores of 2, 3, and 4. For postural tremor, the overall accuracy is 0.913 before dimensionality reduction and 0.782 after. The grading accuracies for the two bradykinesia tasks are 0.889 and 0.891, with misclassifications occurring only between adjacent scores, because adjacent severity levels have no clear boundary and transition gradually.
7. Practical Guidance for DAISCH IMU Users
Putting the method into engineering practice, a 6-axis IMU is fully sufficient; four steps matter most. First, confirm range and noise during selection: the angular velocity peak of fine upper-limb movement may approach 1000°/s, and a ±8 g accelerometer range covers most test tasks; for static tremor monitoring only, a lower-range, lower-noise module can be used. Second, unify the sampling rate at 100 Hz and give all nodes a common timestamp or hardware synchronization pulse; without aligned multi-node data, cycle and pause-time calculations will be wrong. Third, separate tremor before calculating bradykinesia metrics; otherwise tremor contaminates RMS, peak, and cycle statistics. Fourth, treat the score as a relative trend rather than an absolute diagnosis; changes across repeated measurements in the same patient are more valuable than a single score.
DAISCH’s IM7 and IM8 series 6-axis IMU sensor modules use a 3-axis accelerometer plus a 3-axis gyroscope for six-degree-of-freedom measurement and contain no magnetometer; this kind of upper-limb relative-motion quantification does not depend on magnetic heading. With multi-node synchronized acquisition and host-computer data logging, the complete chain from raw waveform to severity grading can be reproduced.